For 35+ years A. P. BUCK, INC. has manufactured the most advanced and innovative instruments for indoor air sampling.We supply air sampling pumps and calibrators for industrial hygiene, indoor air quality, occupational health, environmental, health physics and safety disciplines. For more information on our product lines just click on the link that interests you.

                                               

Tetracholoethylene (Cl2C:CCl2)
Item Item Details
(Price is for domestic US market only)
Tetracholoethylene (Cl2C:CCl2)
Item Name:
Tetracholoethylene (Cl2C:CCl2)
Item #:
GAS-133LL

Product Image Gallery
 
Tetrachloroethylene Cl2C:CCl2 NO.133LL


Performance

Measuring Range

0.1 to 0.2ppm

0.2 to 3.0ppm

3.0 to 9.0ppm

Number of Pump Strokes

2

1

1/2

Correction Factor

1/2

1

3

Sampling Time

1.5 minute per pump stroke

Detecting Limit

0.05 ppm (n=2)

Color Change

Yellow Purple

Reaction Principle

Tetrachloroethylene is decomposed by nascent oxygen by oxidizing agent
to liberate hydrogen chloride which discolors indicator to purple.

Coefficient of Variation

10% (for 0.2 to 1 ppm), 5% (for 1 to 3 ppm)

Shelf Life 2 Years
Corrections for
temperature & humidity
Temperature correction is necessary
Store the tubes in the refrigerator to keep at 10oC (50oF) or below.


Possible coexisting substances and their interferences (NOTE)

Substance

Concentration

Interference

Change color by itself

Hydrogen chloride, Chlorine,

1/2 time

Plus error

Produce purple stain

1,2 Dichloroethylene

-

Plus error

Produce purple stain

1,1,1-Trichloroethane

80 ppm

No effect

No discoloration up to 80 ppm

Toluene, Xylene

-

No effect

No discoloration


Calibration gas generation Diffusion tube method


TLV-TWA TLV-STEL Explosive range
25ppm 100ppm -


 
Range 0.1-9


Money Orders and personal Checks also Accepted, choose Pay By Mail during checkout!

Notice: This publication is intended for general information only and should not be used as a substitute for reviewing applicable government regulations, equipment operating instructions, or legal standards. The information contained in this document should not be construed as legal advice or opinion nor as a final authority on legal or regulatory procedures. A.P. BUCK Inc reserves the right to make changes to information, features, and specifications of any product without notice. A.P. BUCK Inc has made every effort to ensure the accuracy of information displayed on this Website, however, A.P. BUCK Inc. is not responsible for typographical errors appearing on this Website. Price and availability information is subject to change without notice.

View A.P.BUCK INC. Privacy Policy and Legal Notice
A. P. Buck Inc.'s Quality Policy is in Compliance with IS0 9002-1994
©1999-2022 A.P. BUCK INC.

A.P. Buck, Inc